Cable connections

EMC measures on the bottom plate

Fitting the cables X1, X8 and X9 on the bottom plate

3
Figure 3 Bottom plate
1
Fit the bottom plate at the bottom of the generator stand as shown in Figure 3.
2
Carfully cut away the insulation on cables X1, X8 and X9 to lay bare the shield braid
(1).
3
Select proper clamps (2) with regard to the thickness of the cables, cable clamps in
different sizes are provided on the bottom plate.
4
Fit the clamp (2) over the shield braid and screw tight to ensure satisfactory contact
between shield braid and clamp.
5
Install and tighten the four screws (3) and contact washers (4) at the bottom plate.
NOTE!
Cable X9 shall be clamped to the side wall of the cabinets, if this is not already done see
Figure 7 in this section.
Siemens-Elema AB