Agilent Technologies InfiniiVision 3000 X- Series Advanced Training Manual page 54

Hide thumbs Also See for InfiniiVision 3000 X- Series:
Table of Contents

Advertisement

3
Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs
25
Let's now use the scope's Segmented Memory mode of acquisition to
capture 500 consecutive occurrences of this edge transition violation. But
to complete the rest of this lab your scope must be licensed with the
Segmented Memory option.
Press the
mode.
26
27
28
29
30
31
32
33
Figure 35 Using the scope's Segmented Memory acquisition to selectively capture 500
54
Press the
and
edge violation.
(zoom) front key to turn off the scope's horizontal zoom
Push the horizontal position/delay knob to re- position the trigger point
back to center- screen.
Set the scope's timebase to 100.0 ns/div.
Press the [Run/Stop] front panel to begin repetitive acquisition again.
Press the [Acquire] front panel key.
Press the Segmented softkey; then turn the Entry knob to select 500 as
the number of segments to capture.
Press the Segmented softkey to begin a Segmented Memory acquisition.
Press the Current Segment softkey; then turn the Entry knob to review all
500 captured segments.
Set the Current Segment to 500 and note the time- tag of the last captured
segment.
edge violations.
navigation keys to navigate to each discovered
3000 X-Series Oscilloscopes Advanced Training Guide

Advertisement

Table of Contents
loading

Table of Contents