Your scope's display should now look similar to
scope's default trigger condition (rising edge of channel- 1). Let's now set
up the scope to trigger on one of the wider digital pulses on channel- 2. To
do this we will qualify triggering on the channel- 1 pulse, then delay
arming of the trigger past the 5 MHz analog burst, and then finally trigger
on the Nth occurrence of one of the wider digital pulses.
12
13
14
Figure 45 Using Edge then Edge triggering to synchronize acquisitions on the 3rd digital
Your scope's triggering should now be synchronized on the 3rd digital
pulse on channel- 2 after qualifying on the channel- 1 pulse, and after
delaying past the higher frequency analog burst as shown in
3000 X-Series Oscilloscopes Advanced Training Guide
Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs
Press the [Trigger] front panel key; then select Edge then Edge using the
Entry knob.
Press the Delay softkey; then enter 2.0 µs using the Entry knob.
Press the Nth Edge B softkey; then enter 3 using the Entry knob.
pulse of channel-2.
Figure 44
while using the
Figure