Agilent Technologies InfiniiVision 3000 X- Series Advanced Training Manual page 59

Hide thumbs Also See for InfiniiVision 3000 X- Series:
Table of Contents

Advertisement

Figure 39 Automatic "Search" finds and marks multiple runt pulses.
Your scope's display should now look similar to
triangle marks at the top of the display indicate the location of each
positive runt pulse that the scope found. Let's now set up the scope to
automatically navigate to each "found" runt.
20
21
Let's now use the scope's Segmented Memory mode of acquisition to
capture 500 consecutive occurrences of runt pulses of either polarity and
with no specific time qualification. But to complete the rest of this lab
your scope must be licensed with the Segmented Memory option.
22
23
24
25
26
27
28
3000 X-Series Oscilloscopes Advanced Training Guide
Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs
Press the
(zoom) front panel key to turn on the scope's horizontal
zoom mode.
Press the
and
positive runt pulse.
Press the
(zoom) front panel key to turn off the scope's horizontal
zoom mode.
Push the horizontal position/delay knob to re- position the trigger point
back to center- screen.
Set the scope's timebase to 100.0 ns/div.
Press the [Run/Stop] front panel to begin repetitive acquisition again.
Press the [Trigger] front panel key.
Press the "runt polarity" softkey and select the "either polarity" icon.
Press the Qualifier softkey to select None.
navigation keys to navigate to each discovered
Figure
39. The white
3
59