Agilent Technologies InfiniiVision 3000 X- Series Advanced Training Manual page 82

Hide thumbs Also See for InfiniiVision 3000 X- Series:
Table of Contents

Advertisement

4
Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
Figure 57 Triggering on MOSI data = 03
Your scope should now be triggering on MOSI data = 03
shown at center- screen similar to
stream of SPI traffic and then manually search through the captured and
decoded record.
30
31
32
33
34
As you scroll through the data, note that waveforms "track". This means
that the frame that the arrow points to in the lister table corresponds to
the waveforms that are positioned at center- screen. If you want to zoom
in on a particular SPI frame of data, either push the Entry knob, or press
the Zoom to Selection softkey. Let's now perform an automatic search to find
every occurrence of MOSI data = 03
navigate to each of these occurrences.
35
36
37
38
82
Press the [Serial] front panel key.
Press the Lister softkey; then turn the Entry knob to select Serial 1 (SPI).
Set the scope's timebase to 20.00 ms/div.
Press [Run/Stop] to stop repetitive acquisitions.
Press the Scroll Lister softkey; then turn the Entry knob to manually
scroll through the lister table.
Push the horizontal position/delay knob to re- position the trigger point
back to center- screen.
Set the scope's timebase to 20.00 ms/div.
Press the [Search] front panel key.
Press the Search Edge softkey; then select Serial 1 (SPI).
.
HEX
Figure
57. Let's now capture a very long
. We will then automatically
HEX
3000 X-Series Oscilloscopes Advanced Training Guide