Reattach the cover plate (1/Fig. 32).
Secure the cable to the switch board with cable clamps.
Reattach the lower telescope cover (1/Fig. 33).
Reattach the cover (2/Fig. 33) at the Velcro closures.
Reconnect plug X1 to the cover with the brake switches.
Reconnect the protective conductor to the cover with the brake switches.
Remount the cover with the brake switches to the horizontal carriage.

Concluding work

Check the horizontal lift for ease of movement.
Check the angulation for ease of movement.
After completing all service work and reattaching the covers, perform the protective
conductor test in accordance with ARTD-002.731.17.
Perform a function test on the SIREMOBIL.