6
Fig. 10
Remove the screws for the end stop (1/Fig.10).
Connect the cables to both microswitches (refer to S3,S4/Fig.3 and Fig.11).
Reconnect X23.
Place the cover with the brake switches on the C-arm and plug in the test board.
Switch ON the SIREMOBIL.
Release the horizontal lift brake (press the switch, the LED lights).
Move the vertical lifting column up and down until you are able to hammer the end stop
(1/Fig.10) out of the horizontal carriage in the direction of the arrows (arrows/Fig. 10)
without applying pressure to the shaft.
Once the end stop has been removed, the shaft should lie horizontal within the
horizontal carriage and should not be pressing against it.
Switch OFF the SIREMOBIL.
SIREMOBIL Iso-C
Replacing the horizontal lift brake
1
Fig. 11
SPR2-230.076.01
Rev. 01
02/99
S3.Comm
S3.NL
S3.2
Page 6 of 16
TD SD 24
S4.Comm
S4.NL
S4.2