Fig. 28
2
3
Fig. 30
Reattach the back cover (1/Fig. 28).
Reattach the plate (1/Fig. 29 .
Reattach the cable guide (1/Fig. 30). Make sure that the cable bundle is routed properly.
Reattach the cable bundle (2/Fig. 30). When doing this, make sure that the horizontal
brake cable cannot rub against the horizontal carriage. If necessary, press it slightly in
the recess with a flat object.
Place the cable clamp over the bracket (3/Fig. 30).
Reattach the cable bundle with the cable clamp (1/Fig. 31). When doing this, make sure
that the cable bundle is routed correctly. (Refer to Fig. 30).
Do not attach the cable for the brake switches (connector X1) with the other cables in the
cable clamp (1/Fig. 31).
SIREMOBIL Iso-C
Replacing the horizontal lift brake
1
1
5
4
Fig. 29
Fig. 31