Linearize (Lin) - Teledyne 2120XL Operating Instructions Manual

Trace nitrogen in argon analyzer
Table of Contents

Advertisement

Operation
Repeat the process for adjusting the 20 mA output. Pressing
ENTER will accept the value and return the display to the last System
screen.
4.5.1.12 L
(L
INEARIZE
With this function, the user has the capability of linearizing the
output of the instrument by flowing prepared samples with known
concentration across the analytical range. The linearization can be
performed manually or automatically depending on the setting chosen in
the linearization setup screen. Upon completion of the linearization
sample introduction, the software performs linear approximation
between data points.
The Model 2120XL can accept up to 11 linearization points. These
are:
Zero point (pure argon)
 9 data points between zero and full span
 Full Span (85-100 ppm N
full range)
The linearization can be performed manually or automatically
depending on the chosen setting.
To enter the linearization function:
Press the SYSTEM key to display the first System menu. Use the
◄►arrow keys to position the blinking over the More function, and
press ENTER. Repeat the above to ENTER at More a second time. This
will display a third screen of system functions as shown below.
Use the ◄►arrow keys to position the blinking over the Lin func-
tion, and press ENTER. The following screen will appear:
4.5.1.12.2 Setup Sub Function
The Setup sub function allows the operator to choose between
Manual or Automatic linearization. In manual mode the operator the
operator enters the value of the prepared linearization sample as reported
Teledyne Analytical Instruments
)
IN
in Ar or 85-100% of the highest
2
Model Out-cal Lin
Linearize
Manual
Model 2120XL
Auto
52