Leakage Check
[1]
Fill the DEVELOPER, FIXER, and WASH TANKS to the overflow level with water.
To allow correct operation of the LEVEL SENSING PROBES, check that the level of the solutions
is equal at both the drive and non-drive sides of the DEVELOPER and FIXER TANKS. The
solution must cover the LEVEL SENSING PROBES, located on the non-drive side, at the top
edge of the TANK ASSEMBLY. When the LEVEL SENSING PROBES are not covered with
solution, error codes E032 and E033 can occur and replenishment solution is wasted.
[2]
To provide for conduction of the SOLUTION

LEVEL SENSORS—

(a)
Add 3 ounces of developer to the
DEVELOPER TANK.
(b)
Add 3 ounces of fixer to the FIXER
TANK.
[3]
Move AUXILIARY CIRCUIT BREAKERS CB2,
CB3, and CB4 to the ''I'' position.
[4]
Move the MAIN CIRCUIT BREAKER CB1 to
the ''I'' position.
42
MAIN
CIRCUIT BREAKER
(CB1)
AUXILIARY CIRCUIT BREAKERS
CB2, CB3, and CB4
Figure 49 Location of Circuit Breakers
H108_0002CCC
H108_0320CA
1C0932