1.2    Safety  Considerations
This  device  is  marked  with  the  international  caution  symbol.  It  is  
important  to  read  this  manual  before  installing  or  commissioning  this  
device  as  it  contains  important  information  relating  to  Safety  and  
EMC  (Electromagnetic  Compatibility).
This  device  is  a  panel  mount  device  protected  in  accordance  with    
EN  61010-­1:2001,  electrical  safety  requirements  for  electrical  
equipment  for  measurement,  control  and  laboratory.  Installation  of  this  
operation,  the  following  instructions  should  be  followed.
This  device  has  no  power-­on  switch.   A n  external  switch  or  circuit-­
breaker  shall  be  included  in  the  building  installation  as  a  disconnecting  
device.    It  shall  be  marked  to  indicate  this  function,  and  it  shall  be  in  
close  proximity  to  the  equipment  within  easy  reach  of  the  operator.    
The  switch  or  circuit-­breaker    shall  meet  the  relevant  requirements  of  
IEC  947–1  and  IEC  947-­3  (International  Electrotechnical  Commission).  
The  switch  shall  not  be  incorporated  in  the  main  supply  cord.
Furthermore,  to  provide  protection  against  excessive  energy  being  
drawn  from  the  main  supply  in  case  of  a  fault  in  the  equipment,  an  
overcurrent  protection  device  shall  be  installed.
•    Do  not  exceed  voltage  rating  on  the  label  located  on  the  top  of
     the  device  housing.
•     A lways  disconnect  power  before  changing  signal  and  power
     connections.
•    Do  not  use  this  device  on  a  work  bench  without  its  case  for
     safety  reasons.
•    Unit  mounting  should  allow  for  adequate  ventilation  to  ensure
     device  does  not  exceed  operating  temperature  rating.
•    Use  electrical  wires  with  adequate  size  to  handle  mechanical
     strain  and  power  requirements.  Install  without  exposing  bare  wire
     outside  the  connector  to  minimize  electrical  shock  hazards.
EMC  Considerations
•    Whenever  EMC  is  an  issue,  always  use  shielded  cables.
•    Never  run  signal  and  power  wires  in  the  same  conduit.
•    Use  signal  wire  connections  with  twisted-­pair  cables.
•    Install  Ferrite  Bead(s)  on  signal  wires  close  to  the  device  if
     EMC  problems  persist.
Failure  to  follow  all  instructions  and  warnings  may  result  in  injury!