Two  /  Three  /  Four  Wire  Rtd-­Hookups; Figure  2.6 Two/Three/Four-­Wire  Rtd  Wiring  Hookup A)  Rtd-­1000  Ohm/Rtd-­500  Ohm; B)  Rtd-­100  Ohm - Omega i Series User Manual

Temperature/process limit controller
Hide thumbs Also See for i Series:
Table of Contents

Advertisement

2.3.3    Two/Three/Four-­Wire  RTD
(shown  in  bold  lines)  required  to  hookup  a  2-­,  3-­  or  4-­wire  RTD.
Figure  2.6  a)   RTD-­100  ohm  and    
 
500  ohm  Wiring    
 
Hookup
The  two-­wire  connection  is  simplest  method,  but  does  not  compensate  for  
lead-­wire  temperature  change  and  often  requires  calibration  to  cancel  lead-­wire  
resistance  offset.
The  three-­wire  connection  works  best  with  RTD  leads  closely  equal  in  
resistance.  The  device  measures  the  RTD,  plus  upper  and  lower  lead  drop  
voltage  and  the  subtracts  twice  the  measured  drop  in  the  lower  supply  
current  lead  producing  excellent  lead-­resistance  cancellation  for  balanced  
measurements.
The  four-­wire  RTD  hookup  is  applicable  to  unbalanced  lead  resistance  and  
enables  the  device  to  measure  and  subtract  the  lead  voltage,  which  produces  
the  best  lead-­resistance  cancellation.
Type  menu  (see  Part  3).
If  the  input  wires  of  the  meter  get  disconnected  or  broken,  it  will  display
+OPN
RTD.  In  this  case  the  display  shows  
For  safety  purpose  you  may  want  to  set  up  your  alarm  to  be  triggered
when  input  is  open.  See  Alarm  1  &  2  Sections  3.2.6,  3.2.7  for  details.
RTD (1000/500) 4-Wire
RTD (1000/500) 3-Wire
RTD (1000/500) 2-Wire
b)  RTD-­100  ohm  Wiring  Hookup
 "Input  (-­)  Open"  message.
-‐OPN
8
RTD (100) 4-Wire
RTD (100) 3-Wire
RTD (100) 2-Wire

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cni8-alCni16-alCni32-al

Table of Contents