Part 4: Specifications - Omega i Series User Manual

Temperature & humidity monitor
Hide thumbs Also See for i Series:
Table of Contents

Advertisement

PART  4        SPECIFICATIONS
SENSOR  SPECIFICATIONS
Relative  Humidity  (RH)
Accuracy/Range:  
Non-­linearity:  ±3%
Hysteresis:  ±1%  RH
Response  Time:
4  sec  (63%  slowly  moving  air)
Repeatability:  ±0.1%
Resolution:  0.03%,  12bit
NOTE: Reconditioning  of  the  probe  may
be  necessary  if  the  probe  is  stored  for  a
period  of  time  in  a  harsh  environment
(e.g.  high  humidity  or  exposure  to
chemicals).  To  recondition  the  probe:
heat  probe  for  1  day  at  100°C  to  return
it  to  calibration  conditions.
Temperature  (T)
Accuracy/Range*:
±1°C  (±2°F)  for
-­40  to  0°C  and  80  to  123.8°C
(-­40  to  32°F  and  176  to  254°F)
±0.5°C  (±1°F)  for  0  to  80°C  (32  to  176°F)
*NOTE: extended  temperature  range  is
for  Probe  only,  the  Controller's
operating  temperature  is  0-­50°C
Response  Time:  
5  sec  (63%  slowly  moving  air)
Repeatability:  ±0.1°C
Resolution:  0.01°C,  14  bit
METER  SPECIFICATIONS
NMRR:    60  dB
CMRR:    120  dB
A/D  Conversion:
12  bit  RH  and  14  bit  Temperature
Reading  Rate:  
2  samples  per  seconds  max.
Digital  Filter:    Programmable
Decimal  Selection:
None,  0.1  for  temp  and  humidity
Warm  up  to  Rated  Accuracy:  30  min.
Display:    4-­digit,  9-­segment  LED
10.2  mm  (0.40"):  i32,  i16D,  i8DV
10.2  mm  (0.40")  &  21  mm  (0.83"):i8DH
red,  green  and  amber  programmable
colors  for  process  variable,  set  point
and  temperature  units
ALARM  1  &  2:      (programmable  to
display  color  change)
Relay*:    250  Vac  or  30  Vdc  @  3   A
(Resistive  Load);;  configurable  for
on/off,  PID  and  Ramp  and  Soak
Output  1*: SPDT type,  can  be
configured  as   A larm  1  output
Output  2*: SPDT type,  can  be
configured  as   A larm  2  output
SSR*:    20-­265  Vac  @  0.05-­0.5   A
(Resistive  Load);;  continuous
DC  Pulse*:    Non-­Isolated;;  
10  Vdc  @  20  mA
Operation:   high/low,  above/below,
band,  latch/unlatch,  normally
open/normally  closed  and
process/deviation;;  front  panel
configurations  
*  Only  with  -­AL Limit   A larm  Option
NETWORK   A ND  COMMUNICATIONS
(Optional  -­C24,  -­C4EI,  -­EI)
Ethernet:  Standards  Compliance
IEEE  802.3  10Base-­T
Supported  Protocols: TCP/IP,   A RP,
HTTPGET
RS-­232/RS-­422/RS-­485/MODBUS:
Selectable  from  menu;;  both   A SCII
and  modbus  protocol  selectable  from
menu.  
Programmable  300  to  19.2  K  baud;;
complete  programmable  setup
capability;;  program  to  transmit  current
display,  alarm  status,  min/max,  actual
measured  input  value  and  status.
RS-­485:  Addressable  from  0  to  199
Connection:  Screw    terminals
34