Specifications - Omega PHCN-37 User Manual

Microprocessor-based ph controller and ph/orp controller
Hide thumbs Also See for PHCN-37:
Table of Contents

Advertisement

INPUT SIGNAL
pH and Volt Specification
Range:
Resolution:
Accuracy:
Calibration:
Temperature Compensation:
Volt specification is only used to verify proper operation of electrodes.
ORP Specification
Range:
Resolution:
Accuracy:
ORP specification is only used to verify proper operation of electrodes.
TEMPERATURE
Range:
Resolution:
Accuracy:
Unit of Measure:
Sensor:
Isolation:
Noise Rejection:
Display:
Symbol:
Input Impedance:
SECTION 9. SPECIFICATIONS
- 2.00 to 16.00 pH (-620 to +620 mV)
1, 0.1, 0.01 pH
± 0.01 pH (± 0.1 mV)
Two or Three Point
Automatic: 0ºC to 100ºC (32ºF to 212ºF)
Manual: 0ºC to 100ºC (32ºF to 212º F)
-620 to +620 mV
From: -199.9 to 620.0 mV: 0.1 mV
From: -200 to -620: 1 mV
± 0.1 mV
0ºC to 100ºC (32ºF to 212ºF)
0.1ºC
±0.5ºC
ºF or ºC (programmable by software)
100/1000 Ohm Pt RTD (Connection by screw terminals)
Dielectric strength to 2500V transient per 3mm spacing
based on EN 61010 for 260Vrms or dc working voltage
Normal Mode Rejection (NMR) = 60dB
Common Mode Rejection (CMR) = 120dB
LED 14-segment, 13.8 mm (0.54")
>10
12
ohm

Specifications

(
From: -199.9 to 620.0 mV: 0.1 mV
From: -200 to -620: 1 mV
9
)
41