Additional Functions - Miele PG 8504 Operating Instructions Manual

Commercial dishwasher for laboratory glassware and utensils
Table of Contents

Advertisement

Additional functions

You can use this menu to customise the current programme to suit
technical requirements and the load items or to reset all additional
functions to the factory default settings.
Additional specialist knowledge is required to alter programme
settings and this should therefore be undertaken only by experienced
users or by the Miele Customer Service Department.
Changing programme parameters on a validated machine will
necessitate a renewed performance validation.
Further settings
Open the
the  button and then switch it on again with the  button whilst
keeping the  button held in.
Open the menu as follows:
     
Further settings
Additional functions
         
Reset
-
All parameters that have been set under "Additional functions" will
be reset to the factory setting.
Increased water level
-
The water level will be increased for all programmes.
Interim rinse
-
All programmes that this option is intended for will have an extra
interim rinse stage (see the programme chart).
-
Dispensing system
Vent the dispensing system and rename it.
-
Temperature / time
Adjust the temperature and holding time for the Main wash or Final
rinse programme block.
Use the  and  arrow buttons to select an option and confirm
your selection with OK.
See the next section for details of how to continue.
menu by switching the machine off with
Additional functions
Reset
Increased water level
Further settings
77

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pg 8504 adPg 8504 adp

Table of Contents