Areas of application
40
Carry out a visual check before starting every programme:
- Is everything correctly loaded/connected for cleaning?
- Was the recommended loading template followed?
- Can the lumen / narrow sections of hollow items be accessed by
the wash fluid?
- Are the spray arms clean and do they rotate freely?
- Are the filters clean?
Remove any coarse soiling and clean them if necessary.
- Are the removable modules, injector nozzles, irrigation sleeves and
other rinsing fittings securely connected?
- Are the baskets and modules or mobile units correctly connected to
the water supply and are the water connectors undamaged?
- Are all chemical containers sufficiently filled?
The following must be checked at the end of every programme:
- Carry out a visual check of the load for cleanliness.
- Check that all hollow items are still securely located on their injector
nozzles.
 Any hollow items that have become disconnected from their
fittings during reprocessing must be reprocessed.
- Check that the lumen of hollow items are free of obstruction.
- Check that injector nozzles and connectors are securely held in
position in the baskets or inserts.