Diagnostics and troubleshooting
9.4 Treatment of measurement errors
Code
Text message
Error in the delivery
status
M501
Error in the delivery
status
M507
Error in the
transmitter settings
M508
No executable
Bluetooth software
M509
Software update
running
M510
No communication
with the main
controller
M511
Inconsistent software
configuration
9.4

Treatment of measurement errors

The tables below give typical examples of application-related measurement errors.
The images in column " Error description" show the actual level as a dashed line and
the output level as a solid line.
1
2
Note
If the output level is constant, the cause could also be the fault setting of the current
output to " Hold value".
If the level is too low, the reason could be a line resistance that is too high
30
Cause
Hardware error EEPROM
Error during setup
Error when carrying out a reset
Auto false echo suppression faulty
Checksum error in Bluetooth software Carry out software update
Software update running
Communication between main
electronics and display module
disturbed
A software unit requires a software
update
Real level
Level displayed by the transmitter
Rectification
Load XML file with transmitter data
into the transmitter
Send transmitter for repair
Carry out reset and repeat setup
Wait until software update is finished
Check the connection cable to the
display
Send transmitter for repair
Carry out software update
Operating Instructions, 12/2020, A5E49829013