Siemens SITRANS L Operating Instructions Manual page 36

Radar transmitters
Hide thumbs Also See for SITRANS L:
Table of Contents

Advertisement

Code
Text message
F105
Determine measured
value
F260
Error in the calibration
F261
Error in the
transmitter settings
F265
Measurement
function disturbed
Function check
Code
Text message
C700
Simulation active
Out of specification
Code
Text message
S600
Impermissible
electronics
temperature
S601
Overfilling
S603
Impermissible
operating voltage
Maintenance
Code
Text message
M500
SITRANS LR110
Operating Instructions, 12/2020, A5E49829013
Cause
The transmitter is still in the switch-on
phase, the measured value could not
yet be determined
Checksum error in the calibration
values
Error in the EEPROM
Error during setup
Auto false echo suppression faulty
Error when carrying out a reset
Program sequence of the measuring
function disturbed
Cause
A simulation is active
Cause
Temperature of the electronics in the
non-specified range
Danger of vessel overfilling
Terminal voltage too small
Cause
The data could not be restored during
the reset to delivery status
Diagnostics and troubleshooting
9.3 Status messages according to NE 107
Rectification
Wait for the end of the switch-on
phase
Duration up to 3 minutes depending
on the measurement environment
and parameter settings
Send transmitter for repair
Repeat setup
Carry out a reset
Device restarts automatically
Rectification
Finish simulation
Wait for the automatic end after
60 mins.
Rectification
Check ambient temperature
Insulate electronics
Make sure that there is no further
filling
Check level in the vessel
Check terminal voltage, increase
operating voltage
Rectification
Repeat reset
29