Faults; Fault Memory; Troubleshooting; Cleaning The Volume Meter - Bosch KMA 800 Original Instructions Manual

Hide thumbs Also See for KMA 800:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 23
en
34 | KMA 802/822 | Faults
6.

Faults

6.1

fault memory

Any errors occurring in the system are indicated in a
dialog window. In addition, the operator display unit
KMA 800/822 is equipped with a fault storage function.
This is located in the software EPS945-PE/VE/CR on the
program path Configuration / System / Status Informa-
tion or in the measurement display under menu Proto-
col / Status Information. Here, the latest error reports
are filed. Each time the software EPS945-PE/VE/CR is
restarted, all error reports are erased. Error specificati-
on is carried out by error codes.
6.2

Troubleshooting

Fault
Remedy
No display on operator
Check that monitor and PC are con-
and display device.
nected and switched on.
With the KMA 822, check that the so-
cket strip on the trolley is connected
and switched on.
No power to EPS socket strip (refer
to instructions for basic test bench).
Check that VGA connecting cable is
connected to PC and to monitor.
When channel is detec-
Measuring system is not yet comple-
ted, individual connec-
tely filled with test oil (initial start-
ted channels are not de-
up). Run channel detection several
tected.
times in succession, or set manually
on an adapted channel.
During flow measure-
Volume meter contaminated. (volume
ment, severe fluctua-
meter B1 on angled channels and vo-
tions of straight and
lume meter B2 on straight channels).
angled channels. Oil le-
This is detected by the extent that
vel in supply hose of vo-
the oil level rises in the supply hose
lume meter rises and
to the volume meter.
falls during the test.
Clean and bleed volume meter (see
Chapter 6.3).
Straight or angled chan-
Volume meter jammed or defective
nels show no volume.
(volume meter B1 on angled chan-
Supply hose to the volu-
nels and volume meter B2 on straight
me meter is completely
channels). Detection as above.
filled with oil.
Clean and bleed volume meter (see
Chapter 6.3).
|
1 689 979 674
2013-12-23
6.3

Cleaning the volume meter

6.3.1
Cleaning the volume meter
Always wear protective goggles during as-
sembly work in the hydraulics area of the
KMA 802/822.
Always wear protective gloves when working
on the KMA 802/822.
!
Inexpert disassembly and assembly may impair ope-
ration or the accuracy of the volume meter (fig. 12,
item 1 and 2).
i
Complete removal of the volume meter in order to
clean the control piston or the gear pump is not
necessary.
i
Particles in gear pump:
No measurement values, as pump is blocked. Supply
hose to the volume meter is completely filled with
oil.
i
Particles in control piston:
Bar charts and measurement values fluctuate con-
siderably. Oil level in the supply hose to the volume
meter rises and falls.
B1
1
Fig. 12:
Electronics area of measuring device
B2
458737-25_Pal
2
Robert Bosch GmbH