FREQ. RESPONSE Menu Item
This sets the frequency response, step response, and resolution
of the channel according to the table below. The tracking value,
peak/valley detector values and analog output on this channel
are all affected. The default value of 016. HERTZ is suitable for
most applications.
Frequency
Step
Response
Response
(ms) (typical)
40
002. HERTZ/
FAST
440
002. HERTZ
110
008. HERTZ
55
016. HERTZ
28
032. HERTZ
16
050. HERTZ
8
100. HERTZ
3
250. HERTZ
2
500. HERTZ
2
800. HERTZ
POWER-ON SOURCE Menu Item
This menu selects which value is displayed by the channel when
first entering the RUN mode.
The choices are:
"TRACK" means the live tracking value of the channel.
"PEAK" means the highest value of the channel.
"VALLEY" means the lowest value of the channel.
"TRACK / PEAK" means the live tracking value and the highest
value of the channel. (Only available on Model SC3004).
"TRACK / VALLEY" means the live tracking value and the low-
est value of the channel. (Only available on Model SC3004).
"PEAK / VALLEY" means the highest and lowest values of the
channel. (Only available on Model SC3004).
58
Honeywell Sensing and Control
Resolution (counts) (not
including min. 10% overrange/
underrange capability)
5V & 10V
1V & 4-20mA
Inputs
Inputs
±50000
±50000
±50000
±50000
±25000
±25000
±25000
±20000
±20000
±10000
±15000
±10000
±10000
±5000
±10000
±5000
±4000
±2500
±2500
±2500

12.6.2 DISPLAY SETUP Sub-Menu

This menu controls how data values are displayed by the channel
and transmitted via serial communications.
DISPLAY. DIGITS Menu Item
Selects the number of digits displayed by the channel. The
choices are:
"5 DIGIT-BIPOLAR" displays both positive and negative num-
bers with five full digits (±99999).
"6 DIGIT-UNIPLOAR" displays positive numbers with six full
digits (999999) and negative numbers with five full digits
(-99999).
"7 DIGIT UNIPOLAR" displays positive number with seven
full digits (9999999) and negative numbers with six digits
(-999999).
DISPLAY. DECPT Menu Item
Selects the decimal point location on the channel's display and
serial communications output. Use the [UP] and [DOWN] buttons
to move the decimal point to the right and left.
This menu item may be automatically updated by a transducer's
Signature Module.
DISPLAY. COUNT-BY Menu Item
Determines by what value the display increments or decrements.
This will make the display less sensitive to a noisy signal. The
choices are:
"00001"
"00002"
"00005"
"00010"
"00020"
"00050"
"00100"
"00200"
DISPLAY. UNITS Menu Item
Specifies the four character label that is displayed to the right of
the channel's values.
NOTICE
This menu item doesn't change the mathematical scaling of
the channel's values; that can be changed by altering the
"CALIBRATION DATA -> FULL SCALE VALUE" menu item and then
performing a re-calibration.
This menu item doesn't change the mathematical scaling of the
channel's values; that can be changed by altering the "CALIBRA-
TION DATA -> FULL SCALE VALUE" menu item and then performing
a re-calibration.
When a character position is flashing press the [UP] or [DOWN]
button to change the character. Press [ENTER] to advance to the
next character.