Chapter 11 Ac/Ac-Lvdt Input Channel; Features; Wiring; Table 11-1: Ac/Ac-Lvdt Input Channel Pin Connections - Honeywell SC Series Installation Instructions Manual

Signal conditioning self-calibrating digital indicators
Table of Contents

Advertisement

SC 2000, SC2001, SC3004
Chapter 11 AC/AC-LVDT Input
Channel

11.1 Features

The ac/ac-LVDT Input channel provides an ac excitation voltage
to and accepts ac signals from LVDT (Linear Variable Differential
Transformer) transducers. These signals are digitized, converted
into engineering units, and placed into the track, peak and valley
data values of the channel. Setup and calibration of the channel
are made manually through the SETUP mode.
The analog-to-digital converter features adjustable digital, low-
pass filtering, 12- to 18-bit resolution (depending on the filter
setting) and has several different input ranges. These many input
ranges allow ±25,000 count resolution (at the slowest filter set-
ting) for a wide variety of LVDT outputs.
The ac/ac-LVDT Input channel is calibrated to the transducer by
using known-displacement calibration.
Two rear panel control inputs can be field-configured for such
functions as remote tare, disabling peak/valley detection and
clearing the peak/valley values. A standard voltage or optional
current digital-to-analog output is also provided.
NOTICE
To use a dc/dc LVDT with an SC instrument, use a High-Level
Input channel instead of an ac/ac-LVDT Input channel.
Figure 11-1: Connection of Four- or Five-wire ac/ac-LVDT
AC-AC LVDT TRANSDUCER
(+)SUPPLY
(+)OUTPUT
N/C
OUTPUT RETURN
SUPPLY RETURN

11.2 Wiring

Connect your transducer to an ac/ac-LVDT Input channel by
wiring it to the 12-pin connector of that channel. The Customer In-
formation Sheet that shipped with the instrument describes which
cards are installed in each channel. The pin-out for this connector
is shown on the following table.

Table 11-1: ac/ac-LVDT Input Channel Pin Connections

Pin
Label
Function
1 (top)
+EXC
(+)Excitation
2
N/C
No connection
3
N/C
No connection
4
-EXC
(-)Excitation
5
+SIG
(+)Signal
6
-SIG
(-)Signal
7
+OUT
Analog Output
8
-OUT
Analog Return
9
N/C
No connection
10
DGND
Digital Ground
11
AUX1
Auxiliary Function 1 (connect
to pin 10 to activate)
11 (bot-
AUX2
Auxiliary Function 2 (connect
tom)
to pin 10 to activate)
The Analog Output and Analog Return pins are electrically iso-
lated from all other pins on the instrument.
CABLE
008-0608-00
Rev. G,
Reference
Pin
10
10
10
10
8
-
-
10
10
INSTRUMENT CONNECTIONS
DESIGNATION
(+)EXCITATION
N/C
N/C
(-)EXCITATION
(+)SIGNAL
(-)SIGNAL
(+)ANALOG OUTPUT
(-)ANALOG OUTPUT
N/C
DGND
AUX1
AUX2
NOTE KEYED CONNECTOR
CABLE SHIELD CONNECTION SCREW
(CONNECT TO CABLE SHIELD)
Honeywell Sensing and Control
PIN
1
2
3
4
5
6
7
8
9
10
11
12
39