Notch Filter Rejection
SPECIFICATION
Rejection (995 to 1025 Hz); >50 dB
DESCRIPTION
An AC Calibrator is used to check the HP 4934A Notch Filter rejection.
EQUIPMENT
AC Calibrator
: DATRON 4707A
PROCEDURE
CAUTION
Ensure that the AC Calibrator OUTPUT is switched off or set to the minimum level
_ before proceeding with the following procedure.
1
Connect the AC Calibrator output to the HP 4934A RCV port.
2
Set the HP 4934A as follows:
Measurement
: NOISE W TONE
Filter
: C-MSG
Receiver (RCV) Impedance
: 600Q
Display Mode
: RCV
3
Set the AC Calibrator output level to 0.7746 V at 995 Hz then switch on the
OUTPUT.
4
Check the reading on the HP 4934A is less than 40 dBm.
5
Set the AC Calibrator to each of the frequencies in the following table and check
the HP 4934A noise readings.
AC Calibrator Frequency
HP 4934A Noise Reading
1004 Hz
<40 dBm
1015 Hz
<40 dBm
1025 Hz
<40 dBm
5-36
Performance Tests