HP 4934A Operating And Calibration Manual page 78

Transmission impairment measuring set
Table of Contents

Advertisement

Hewlett-Packard
Tested by:
Model 4934A
Date:
Transmission Impairment Measuring Set
Serial No.:
Table 1
Operation Verification Test Record
Page No.
Test Description
Result
Min
Actual
Max
Transmitter Frequency Accuracy
549
Step 3:
20 Hz
19.999 Hz
20.001 Hz
Step 4:
1000 Hz
999.95 Hz
1000.05 Hz
50,000 Hz
49,997.5 Hz
50,002.5 Hz
110,000 Hz
109,986.8 Hz
110,013.2 Hz
Transmitter Level Accuracy
5-21
Step 4:
OdBm
0.7657 V
0.7836 V
Step 5:
-5 dBm
0.4306 V
0.4406 V
-10 dBm
0.2421 V
0.2478 V
-19 dBm
0.0859 V
0.0879 V
Receiver Frequency
5-4
Step 3:
20 Hz
19 Hz
21 Hz
Step 4:
1000 Hz
998 Hz
1002 Hz
50,000 Hz
49.99 kHz
50,01 kHz
110,000 Hz
109.99 kHz
110.01 kHz
Receiver Level
5-5
Step 3:
OdBm
-0.2 dBm
0.2 dBm
Step 5:
-19 dBm
-19.2 dBm
-18.8 dBm
5-14
Performance Tests

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents