Hold Circuit
WARNING
Hazardous voltages are present during this procedure - only service trained
personnel, aware of the hazards involved, should perform the test.
EQUIPMENT
DC Voltmeter
: HP 3456A
DC Power Supply
: HP 6234A
lOOQ 1% Resistor
:HP 0757-0178
WECO 310 Connector
: HP 1252-0695
PROCEDURE
1
Connect the equipment as shown in the following figure.
2
Set the HP 4934A as follows:
Measurement
Transmitter (TRMT) Impedance
Receiver (RCV) Impedance
Transmit Hold Select
Receive Hold Select
LEVEL FREQ
600
n
6000
.
TRMT HOLD
RCV HOLD
Performance Tests
5-11