Table 10.9: Ranges And Accuracy Of Measurements - ResMed Elisee 150 Clinical Manual

Hide thumbs Also See for Elisee 150:
Table of Contents

Advertisement

Measurements

Table 10.9: Ranges and accuracy of measurements

Measurement (unit)
FiO
(%) with 100mL < Vt < 500mL
2
FiO
(%) with 500mL < Vt <1000mL
2
NOT013189-6
PEEP (cmH
O)
2
Pmean (cmH
O)
2
Ppeak (cmH
O)
2
Ftot (bpm)
Ti (s)
Te (s)
I/E
e (L/min)
esp (L/min)
Vte (mL)
Vti (mL)
Range
Accuracy
0 to 100
+/- 5%
0 to 100
+/- 5%
0 to 100
+/- 5%
0 to 100
+/- 10%
0 to 100
+/- 5%
1 to 99
+/- 1 or 5%
0.1 to 9.99
+/- 0.02
10.0 to 99.9
+/- 0.02
0.1 to 9.99
+/- 0.02
1/9.9 to 1/0.1
+/- 0.04
1/99 to 1/9.9
+/- 0.04
0.1 to 99.9
+/- 15%
0.1 to 99.9
+/- 15%
Double circuit
0 to 100
+/- 10%
> 100
+/- 10%
Single circuit
0 to 100
+/- 10%
> 100
+/- 10%
Display accuracy
0.1
0.1
0.1
1
1
1
0.01
0.1
0.01
1/0.1
1/1
0.1
0.1
1
1
1
1