7.4 Alarm checking procedure
Before using the ventilator, check that all critical alarms are properly adjusted according to the
ventilation parameters.
All visual and audible alarms must be tested regularly to ensure they are working properly,
particularly prior to connection to a patient and before continuous ventilation.
Testing the mains power and external DC power alarms
Connect a patient circuit (single or double limb) fitted with a MAQUET test lung to the ventilator
and perform the manual test. Go to the clinical screens (by pressing the padlock) and set up ACV
mode, non-invasive, adult ventilation.
Turn off any alarms which may be triggered (by pressing the Alarm Silence button twice).
Perform the following test for each power supply type (mains, external DC and external battery):
Table 7.6: Power supply tests
Action
Connect power supply.
Disconnect power supply.
Reconnect power supply.
Press the Alarm Silence button.
120/174
Visual alarm
Off
Flashing
Steady
Off
Audible alarm
Off
Sounding
Off
Off
NOT013189-6