Ranges of adjustment, accuracies and default values

Ventilation parameters

Table 10.5: Range of adjustment for ventilation parameters

Ventilation type
P insp (cmH
PEEP (cmH
Max. flow (L/min)
Rise time (slope)
Recruitment period
Recruitment size (s)
Recruitment pressure
*: If a beep sounds, it means that the I/E ratio has been inverted during adjustment of the
parameters.
152/174
Default values
Value
Patient type
Paediatric
F (bpm)
25
Ti (s)
0.6
I/E*
1/3.0
Ti Max (s)
1.2
Tapnea (s)
10
O)
12
2
PS (cmH
O)
12
2
O)
0
2
V
(mL)
100
T
10
max
3
Flow shape
1
no
(min)
2.0
15
(cmH
O)
2
V
(mL)
no
TS
Fmini (bpm)
25
Range of adjustment
Minimum
Adult
Paediatric
Adult
15
2
2
1.0
0.3
1/2.0
1/0.4
1.0
15
5
10
20
3
5
15
3
5
0
50
300
500
1000
50
5
10
1
1
1 – no
0.2
0.3
3
3
5
no -50
no -300
no
1000
15
no -3
no -6
Incre-
Maximum
ment
Paediatric
Adult
80
50
1
3.0
0.1
1/9.9
1/0.1
3.0
0.1
20
60
1
60
40
60
1
60
40
60
1
20
25
1
500
1000
10
2500
100
40
120
1
4
1
5
1
60
1
3
3
0.1
40
40
1
60
1
500
1000
10
2500
100
80
50
1
NOT013189-6
Accuracy
+/- 1
+/- 0.1
+/- 0.1
+/- 5%
+/- 5%
+/- 5%
10 mL or
+/- 10%
1/F
+/- 0.1
+/- 5%
10 mL or
+/- 10%
+/-1