5.2 Manual test

A manual test is always required before the device is programmed for a new patient or the patient
circuit is modified. To perform a manual test, go to the Main clinical screen (see "Accessing the
clinical screens" on page 63).
TEST
TEST
Figure 5.3: Accessing the manual test.
the 12/06/07 at 14 : 50 : 35
Ctot = 1.2 ml/cmH2O
Ri = 0.2 cmH2O (30 l/min)
Ri = 0.2 cmH2O (60 l/min)
Re = 0.2 cmH2O (30 l/min)
Re = 0.2 cmH2O (60 l/min)
Figure 5.4: Main screen with manual test results.
The results of the last test are displayed. You can choose to return to the main clinical screen or
perform a new test.
NOT013189-6
Last test the 18/06/07
Double circuit : PASS
ACV
ACV
P1: P1:
P1: P1:
P2: P2:
NEW
NEW
PATIENT
PATIENT
Last test
Circuit test : PASS
Kind of circuit : DOUBLE
RESTART
RESTART
Press here to display the
test screen
PS
PS
Press to return to the main
screen
Back
Press to restart test
Clinical Configuration and Use