Static Mixer and Filter

The capillary pump is equipped with a static mixer and an inline filter in front of the
EMPV.
The standard static mixer
The standard static mixer has a volume of typically 420 µl. To reduce the delay volume
of the capillary pump you can remove the mixer.
Conditions to remove the static mixer:
• The delay volume of the capillary pump should be reduced to a minimum for fastest
• The detector is used at medium or low sensitivity.
NOTE
Removing the mixer will result in an increase of the composition wander and higher
detector noise.
1100 Series Capillary LC System Manual
Primary flow overview for standard pump configuration
0 bar
System
pressure
200
500
800
gradient response.
37
100 bar
200 bar
System
System
pressure
pressure
225
250
570
640
995
1190
Optimizing Performance 2
300 bar
400 bar
System
System
pressure
pressure
275
300
710
780
385
1580