Download Print this page

National Instruments SCXI-1121 Calibration Procedure page 7

Hide thumbs Also See for SCXI-1121:

Advertisement

© National Instruments Corporation
Table 1. Completion Network Jumpers
Channel
Jumpers
0
W1 and W2
1
W17 and
W18
2
W27 and
W28
3
W39 and
W40
Configuring the Gain Jumpers
Each input channel has two user-configurable gain stages. The first-stage
gain provides gains of 1, 10, 50, and 100. The second-stage gain provides
gains of 1, 2, 5, 10, and 20. The device ships with the first-stage gain set to
100 (position A) and the second-stage gain set to 10 (position D).
To change the gain setting of a specified channel on the module, refer to
Table 2 for the jumper reference designator of the channel gain stage you
want to change, and Figure 3 for the location of this jumper. Move the
jumper to the position indicated in Table 3.
Enable
Completion
W2
A
B
W1
1
2
3
W18
A
B
W17
1
2
3
W28
A
B
W27
1
2
3
W40
A
B
W39
1
2
3
7
Disable Network
(Factory Setting)
W2
A
B
W1
1
2
3
W18
A
B
W17
1
2
3
W28
A
B
W27
1
2
3
W40
A
B
W39
1
2
3
SCXI-1121 Calibration Procedure

Advertisement

loading