SCXI-1121 Calibration Procedure
Configuring the Excitation Level
Each excitation channel of the SCXI-1121 module has two different current
or voltage levels. You can set a given channel to one of the following level
modes:
In the current mode—0.150 or 0.450 mA
In the voltage mode—3.333 or 10 V
After selecting the excitation mode of operation desired—voltage or
current—as described in the previous section, refer to Table 6 to set the
SCXI-1121 module for the level of operation. The SCXI-1121 module
ships with the voltage mode set to 3.333 V.
Table 6. Excitation Level Jumper Settings
Excitation
Channel
Jumpers
0
W16 and
W26
1
W24 and
W25
2
W36 and
W37
3
W48 and
W49
3.333 V or
0.150 mA
(Factory Setting)
W16
1
2
1
2
3
3
W24
1
2
3
1
W36
1
2
3
1
W48
1
2
3
1
10
10 V or 0.450 mA
W26
W16
1
2
1
2
3
3
W24
W25
1
2
3
1
2
3
W36
W37
1
2
3
1
2
3
W48
W49
1
2
3
1
2
3
W26
W25
2
3
W37
2
3
W49
2
3