Download Print this page

National Instruments SCXI-1121 Calibration Procedure page 20

Hide thumbs Also See for SCXI-1121:

Advertisement

Gain
2
2
5
5
5
10
10
10
20
20
20
50
50
50
100
100
100
200
200
200
250
250
250
500
500
500
SCXI-1121 Calibration Procedure
Table 12. SCXI-1121 Specifications (Continued)
Test Point (V)
Limit (V)
0.0000
0.005141
–2.2500
–4.465708
0.9000
4.534293
0.0000
0.005142
–0.9000
–4.465707
0.4500
4.534387
0.0000
0.005236
–0.4500
–4.465613
0.2250
4.534456
0.0000
0.005305
–0.2250
–4.465544
0.0900
4.534694
0.0000
0.005543
–0.0900
–4.465306
0.0450
4.535095
0.0000
0.005944
–0.0450
–4.464905
0.0225
4.535892
0.0000
0.006741
0.0225
–4.464108
0.0180
4.536294
0.0000
0.007143
–0.0180
–4.463706
0.0090
4.538303
0.0000
0.009152
–0.0090
–4.461698
4Hz filter setting
Upper
Lower
Limit (V)
–0.005141
–4.534292
4.465707
–0.005142
–4.534293
4.465613
–0.005236
–4.534387
4.465544
–0.005305
–4.534456
4.465306
–0.005543
–4.534694
4.464905
–0.005944
–4.535095
4.464108
–0.006741
–4.535892
4.463706
–0.007143
–4.536294
4.461698
–0.009152
–4.538303
20
10kHz filter setting
Upper
Lower
Limit (V)
Limit (V)
0.006542
–0.006542
–4.464307
–4.535693
4.535706
4.464294
0.006555
–0.006555
–4.464294
–4.535706
4.535771
4.464229
0.006620
–0.006620
–4.464229
–4.535771
4.535979
4.464021
0.006828
–0.006828
–4.464021
–4.535979
4.536146
4.463854
0.006995
–0.006995
–4.463854
–4.536146
4.536551
4.463449
0.007400
–0.007400
–4.463449
–4.536551
4.537797
4.462203
0.008646
–0.008646
–4.462203
–4.537797
4.538614
4.461387
0.009463
–0.009463
–4.461387
–4.538614
4.540951
4.459049
0.011800
–0.011800
–4.459049
–4.540951
ni.com