Download Print this page

National Instruments SCXI-1121 Calibration Procedure page 21

Hide thumbs Also See for SCXI-1121:

Advertisement

Gain
1000
1000
1000
2000
2000
2000
1
Value available only when used with the SCXI-1327 high-voltage terminal block
© National Instruments Corporation
Table 12. SCXI-1121 Specifications (Continued)
Test Point (V)
Limit (V)
0.0045
4.542321
0.0000
0.013170
–0.0045
–4.457679
0.00225
4.551389
0.00000
0.022238
–0.00225
–4.448611
4Hz filter setting
Upper
Lower
Limit (V)
4.457679
–0.013170
–4.542321
4.448611
–0.022238
–4.551389
21
10kHz filter setting
Upper
Lower
Limit (V)
Limit (V)
4.546501
4.453499
0.017350
–0.017350
–4.453499
–4.546501
4.558631
4.441369
0.029480
–0.029480
–4.441369
–4.558631
SCXI-1121 Calibration Procedure