Model 3000 TA-XL-EU
The instrument automatically returns to the Analyze mode at the conclusion of the auto span
cycle..
Manual Mode Spanning:
Press Span to start the Span function.
•
Use the ▲▼ keys to toggle between
•
appears, blinking, on the display. Press Enter to move to the next screen.
Press Enter key to continue.
•
Press ▲ (<UP>) to permit modification (
•
Use the arrow keys to enter the oxygen concentration of the span gas you are using
•
(209000.00 if you are using air). The ◄► arrows choose the digit, and the ▲▼
arrows choose the value of the digit.
Press Enter to enter the span value into the system and begin the span calibration.
•
When the Span value displayed on the screen is sufficiently stable, press Enter and
•
the span reading changes to the correct value.
The instrument automatically returns to the Analyze mode.
Setting Alarm Parameters
If you are using password protection, you will need to enter your password to access the
alarm functions. Follow the instructions above for Entering the Password to enter your
password. Once you have clearance to proceed, you can set the alarms.
####
ppm Span
Slope=####
ppm/s
Span: Settling:MAN
<ENT> For Next
AUTO
Calib. Holding time
Cal hold: 5 min
Span Val: 000008.00
<ENT>Span <UP>Mod #
####
% Span
Slope=#### ppm/s
Teledyne Analytical Instruments
and
span settling. Stop when
MAN
) of span value.
Mod #
Quickstart Guide
MAN
13