Teledyne Trace Oxygen Analyzer 3000TA-XL-EU Quick Start Manual page 16

Teledyne trace oxygen analyzer quick start guide
Hide thumbs Also See for Trace Oxygen Analyzer 3000TA-XL-EU:
Table of Contents

Advertisement

Quickstart Guide
Press the Alarm button on the front panel to enter the Alarm function. Make sure that
AL–1 is blinking.
Using the ◄► arrow keys to move the blinking over to AL–1. Then press Enter.
Five parameters can be changed on this screen:
Value of the alarm setpoint,
to 10,000.00 ppm
Out-of-range direction,
Defeated?
Failsafe?
Latching?
To change any of the alarm parameters, use the ◄► arrow keys to move the blinking
over to the specific parameter.
Use the ▲▼ arrow keys to change the value of the parameter. Holding down the key
speeds up the incrementing or decrementing.
Once the parameters for alarm 1 have been set, press Alarms again, and repeat this
procedure for alarm 2 (
To reset a latched alarm, go to
times. (Toggle it to Y and then back to N.)
—or—
Go to
Ltch–
back to Y.)
Set The Range Function
To set the 3 programmable ranges:
Enter the range function mode by pressing the Range button on the front panel.
AL—1
AL—2
Choose Alarm
AL—1 1000 ppm HI
Dft—N Fs—N Ltch—N
HI
or
(Yes/No)
Dft–Y/N
Fs–Y/N
(Yes/No)
(Yes/No).
Ltch–Y/N
AL–2
).
Dft–
and then press either ▲ two times or ▼ two times. (Toggle it to N and
L—1 M—10
H—100 Mode—AUTO
Teledyne Analytical Instruments
ppm (oxygen); value can be set from 0.00
AL–1 ####
LO
and then press either ▲ two times or ▼ two
Trace Oxygen Analyzer
14

Advertisement

Table of Contents
loading

Table of Contents