(3) CONNECTING THE ENDOSCOPE

CAUTION
- Always turn OFF the processor BEFORE
connecting an endoscope, Connecting the endoscope while the
processor is ON might damage the endoscope.
1. Check to ensure the appropriate light guide adapter is mounted
to the EPK- 700. The adapter, AE-P1 is already attached to the
EPK-700.
-Connecting a video endoscope without a light guide
NOTE
adaptor in place will reduce light output at the endoscope
distal end. Attempting to connect a fiberscope without a light
guide adapter and/or an appropriate light guide sleeve in place
will damage the fiberscope and the EPK-700.
2. Check to ensure that the scope locking lever is open.
3. Connect the scope light guide into the processor.
4. Connect the scope electrical connector into the processor.
5. Close the scope locking lever
NOTE
-If using the fiberscope video adaptor module, make
sure the eyepiece of the fiberscope is properly connected to the
module (use adapters as required). Connect the electrical
connector of the module to the electrical connector of the EPK-
700, lining up the red dots on each.
6. Connect the water bottle Air/ Water connector to the
Air/Water receptacle on the endoscope's umbilical connector.
7. Connect the suction tube of the suction device to the suction
nipple on the umbilical connector of the endoscope.
(4) CONNECTING THE PERIPHERAL
EQUIPMENT
1. Referring to the rear panel diagram, connect a TV monitor to
the EPK-700
2. Connect other required equipment such as hard copy
equipment, VCR etc.
CAUTION
- When used in clinical or residential areas near
radio and TV receiver units, this equipment may be subjected
to radio interference.
- To reduce electromagnetic interference, do NOT keep turning
on the main SWITCH of the Equipment hanging a
videoendoscope on a scope-hanger when not operated.
- To avoid and resolve adverse electromagnetic effects, do
NOT operate this equipment near the RF energy equipment.
CAUTION
: FOR EUROPE
- This equipment is a Class A Medical Equipment (specified
EN55011) and is intended for hospital or health care
districts.
- Use the connection cables as specified below and the keybord
with appropriate CE Marking.
Composite Video Cable (1.5m), RGB Video Cable (1.5m),
Y/C Video Cable (1.5m), Control (1,2) Cable (1.5m),
RS-232C Cable (2.0m)
15