7 Troubleshooting
Error code
E57203
E57204
E57205
92
Problem
ACK timeout dropped
Invalid sender
New before end
Solution
3. Replace i3000.
Error source i3000:
1. Check connections:
* J3009, J2010
If fail => replace the main cable.
2. Replace i2000
3. Replace i3000
Error source i2000 or i3000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable.
* J3009, J2010
If fail => replace the main cable.
2. Replace i3000
3. Replace i2000
4. Replace i4000
Error source i4000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable.
2. Replace i4000
3. Replace i3000.
Software error. Take logs and boot the
device.
Error source i3000:
1. Check connections:
* J3009, J2010
If fail => replace the main cable.
2. Replace i2000
3. Replace i3000
Error source i2000 or i3000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable.
* J3009, J2010
If fail => replace the main cable
2. Replace i3000
3. Replace i2000