KaVo ORTHOPANTOMOGRAPH OP 3D Service Manual page 138

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

8 Part replacement instructions
2. Remove the three screws, one on the right side, one on the left side and one below
the collimator.
3. Detach the extension spring.
4. Lift the collimator off.
5. Fit the new collimator to its place.
6. Attach the new collimator by screwing in the three screws, one on each side of the
collimator and one below it.
7. Reattach the extension spring.
8. Connect the three cables to their connectors to the right places on top of the
collimator.
NOTICE! Check that the number in the cable is the same as in the connector.
9. Calibrate the collimator. See instructions in
and Installation manual, chapter Collimator calibration.
138
Collimator adjustment
ORTHOPANTOMOGRAPH
and refer to User
OP 3D

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents