17. Install the horizontal lasers back.

8.3.6 Replacing the main cable set

NOTICE! You must remove the upper shelf right and left cover and the rotator
back and front covers before this procedure.
NOTICE! Take a picture of the connectors in both ends of the main cable set
before you remove any of them. This helps you to connect the new cable set.
Required tools: Allen key 3mm
1. Detach the cables from the upper shelf end (i3000 power supply board).
2. Detach the grounding cable from the i2000 image capture board.
3. Detach all other cables from the i2000 board.
ORTHOPANTOMOGRAPH
OP 3D
8 Part replacement instructions
125