3. Insert the new card.

8.3.16 Replacing the emergency stop switch with cable

1. Push the lip behind the button to release the connector of the emergency stop
switch.
2. Detach the connector.
3. Unscrew the emergency stop switch button.
4. Insert the new emergency stop switch and connector.
5. Attach the connector.
6. Screw in the emergency stop button.

8.3.17 Replacing the i8000 LED indicator board

NOTICE! You need to remove the carriage left and right cover before this
procedure.