Appendix; Measuring Ranges - Siemens SITRANS F Series Operating Instructions Manual

Electromagnetic flowmeters
Hide thumbs Also See for SITRANS F Series:
Table of Contents

Advertisement

Appendix

B.1

Measuring ranges

MAG 5100 W (7ME6520) MI-001 is verified and labeled at a given Q3 and Q3/Q4 = 1.25 and
Q2/Q1 = 1.6 measuring ranges:
Order code: P11
50 (2")
"R" Q3/Q1
40
Q4 [m
/h]
20
3
Q3 [m
/h]
16
3
Q2 [m
/h]
0.64
3
Q1 [m
/h]
0.4
3
Order code: P12
50 (2")
"R" Q
/Q1
63
3
Q4 [m
/h]
20
3
Q3 [m
/h]
16
3
Q2 [m
/h]
0.41
3
Q1 [m
/h]
0.25
3
Order code: P13
50 (2")
"R" Q
/Q1
80
3
Q4 [m
/h]
20
3
Q3 [m
/h]
16
3
Q2 [m
/h]
0.32
3
Q1 [m
/h]
0.20
3
Order code: P16
50 (2")
"R" Q
/Q1
160
3
Q4 [m
/h]
50
3
Q3 [m
/h]
40
3
Q2 [m
/h]
0.4
3
Q1 [m
/h]
0.25
3
Order code: P17
50 (2")
"R" Q
/Q1
200
3
Q4 [m
/h]
50
3
Q3 [m
/h]
40
3
SITRANS FM MAG 5100 W
Operating Instructions, 01/2020, A5E03063678-AC
65 (2½")
80 (3")
100 (4")
40
40
40
31.25
50
78.75
25
40
63
1.0
1.6
2.52
0.63
1.0
1.58
65 (2½")
80 (3")
100 (4")
63
63
63
31.25
50
78.75
25
40
63
0.63
1.02
1.6
0.40
0.63
1.00
65 (2½")
80 (3")
100 (4")
80
80
80
31.25
50
78.75
25
40
63
0.5
0.8
1.26
0.31
0.50
0.79
65 (2½")
80 (3")
100 (4")
160
160
160
78.75
125
200
63
100
160
0.63
1.0
1.6
0.39
0.63
1.0
65 (2½")
80 (3")
100 (4")
200
200
200
78.75
125
200
63
100
160
125 (5")
150 (6")
200 (8")
40
40
40
125
200
312.5
100
160
250
4.0
6.4
10.0
2.5
4.0
6.25
125 (5")
150 (6")
200 (8")
63
63
63
125
200
312.5
100
160
250
2.54
4.06
6.35
1.59
2.54
3.97
125 (5")
150 (6")
200 (8")
80
80
80
125
200
312.5
100
160
250
2.0
3.2
5.0
1.25
2.00
3.13
125 (5")
150 (6")
200 (8")
160
160
160
312.5
500
787.5
250
400
630
2.5
4.0
6.3
1.56
2.5
3.94
125 (5")
150 (6")
200 (8")
200
200
200
312.5
500
787.5
250
400
630
B
250 (10") 300 (12")
40
40
500
787.5
400
630
16.0
25.2
10.0
15.75
250 (10") 300 (12")
63
63
500
787.5
400
630
10.2
16.0
6.35
10.0
250 (10") 300 (12")
80
80
500
787.5
400
630
8.0
12.6
5.00
7.9
250 (10") 300 (12")
160
160
1250
2000
1000
1600
10.0
16.0
6.3
10.0
250 (10") 300 (12")
200
200
1250
2000
1000
1600
67

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans fm mag 5100 w7me6527me658

Table of Contents