Parameters: Measuring ranges: Setting measuring ranges

126
The full-scale values of the measuring ranges can be set in the
third and fourth lines of this dialog. They must lie within the
factory settings, i.e. if an analyzer is factory-set for a total
range from 0 to 2000 mg/m
within this range. The following definitions also apply:
• The smaller MR must not be greater than the higher MR.
• The following input limits apply to the ranges:
– Lower limit: 0.01 times the smaller MR according to
factory setting (label)
– Upper limit: 1.1 times the higher MR according to factory
setting (label)
In the example:
• Smallest MR 1: 0 to 4 mg/m
• Highest MR 2: 4 to 2200 mg/m
, modifications are only possible
3
3
3