Performance Verification; Required Equipment - Teledyne HVD3102 Operator's Manual

High voltage differential probes
Hide thumbs Also See for HVD3102:
Table of Contents

Advertisement

HVD3102 and HVD3106 High-Voltage Differential Probes

Performance Verification

This procedure can be used to verify the warranted characteristics of the HVD3102/HVD3106
High-Voltage Differential Probes. If the product does not meet specifications, it should be
returned to a Teledyne LeCroy service center. There are no user accessible adjustments, so
there is no adjustment procedure.

Required Equipment

The following table lists the test equipment and accessories (or their equivalents) that are
required for performance verification of the HVD3102/HVD3106 High-Voltage Differential
Probes. This procedure is designed to minimize the number of calibrated test instruments
required. Because the input and output connector types may vary on different brands and
models of test instruments, additional adapters or cables may be required.
Only the parameters listed in boldface in the "Minimum requirements" column must be
calibrated to the accuracy indicated.
Description
Digital Multimeter
Function Generator
BNC Coaxial Cable
Calibration Fixture
Banana Plug Adapter
Insulated Banana Plug
Insulated Banana Couplers (2)
22
Minimum Requirements
DC: 0.1% accuracy
AC: 0.2% accuracy to measure
7 mVrms to 7 Vrms at 70 Hz
Output sine wave of 20 Vpp
Male-to-Male 50 Ω Cable
ProBus Extender Cable
Female BNC-to-Dual Banana
Plug
Insulated BNC-to-Shrouded
Banana Plug
Example Equipment
Agilent 34401A
Fluke 8842A-09
Keithley 2001
Agilent 33120A
Stanford Research DS34
Leader LAG-120B
Pomona 5697-36
Teledyne LeCroy
PROBUS-CF01
Pomona 1269
Mueller BU-00260
Mueller BU-5671-B-12-0
Mueller BU-32601-2 (R)
Mueller BU-32601-6 (B)
923278-00