HVD3102/HVD3106 Test Record
Permission is granted to photocopy this page and record the results of the Performance
Verfication procedure on the copy. File the completed record as required by applicable internal
quality procedures.
The numbers preceding the individual data records correspond to steps in the procedure
requiring the recording of data. Results recorded under "Test Result" are the actual specification
limit check. The test limits are included in all of these steps. Record other measurements and
intermediate calculations that support the limit check under "Intermediate Results".
Serial Number:
Asset / Tracking Number:
Date:
Technician:
Equipment
Oscilloscope
Digital Multimeter
Function Generator 1
1 The function generator used in this Performance Verification Procedure is used for making relative measurements. The
output of the generator is measured with a DMM or oscilloscope in this procedure. Thus, the generator is not required to be
calibrated
Step
Description
5
Function Generator Output Voltage
11
Probe Output Voltage
13
÷500 / ÷1000 Gain Error (test limit ± 1 %)
16
Probe Output Voltage
18
÷50 / ÷100 Gain Error (test limit ± 1 %)
923278-00
Model
Serial Number
Intermediate Data
Operator's Manual
Calibration
Due Date
Test Result
V
V
V
%
%
25