Maintenance

Cleaning

Clean only the exterior surfaces of the probe using a soft cloth or swab dampened with water or
75% isopropyl alcohol solution. Do not use harsh chemicals or abrasive cleansers.
CAUTION. The probe is not water proof. Under no circumstances submerge the probe
or allow moisture to penetrate it.

Calibration Interval

This probe has no adjustments. The recommended calibration interval is one year. A
Performance Verification Procedure is included in this manual.

Service Strategy

The HVD3102/HVD3106 probe utilizes fine-pitch surface mount devices. It is, therefore,
impractical to attempt repair in the field. Defective probes must be returned to a Teledyne LeCroy
service facility for diagnosis and exchange.
CAUTION. Do not remove the covers. Refer all servicing to qualified personnel.
A defective probe under warranty will be replaced with a factory refurbished probe. A probe that
is not under warranty can be exchanged for a factory refurbished probe for a modest fee. You
must return the defective probe in order to receive credit for the probe core.

Replacement Parts

The probe accessories can be ordered as a replacement kit, PK-HV-0001, through your local
sales office, and individual probe accessory tips (pair 1 ea. red/black) can be purchased through
the regional customer care centers listed at the back of this manual.Defective probes can be
replaced on an exchange basis. Replacement probes are factory repaired, inspected, and
calibrated to the same standards as a new product. In order to obtain a replacement probe, you
must return the defective probe. Return the probe to your regional customer care center.
923278-00
Operator's Manual
19