Standard programs
5.3.2

Control charts for data from the AQA Log

48
1. Select Recall Data from the main menu.
2. Press AQA Log.
A listing of the stored data is displayed.
3. Press Filter: On/Off.
4. Activate On.
5. Select a required parameter.
6. Press OK to confirm the selection.
The chosen items are listed.
7. Press Options > Control Chart.