Pc And Printer - Hach DR 6000 User Manual

Laboratory spectrophotometer for water testing
Hide thumbs Also See for DR 6000:
Table of Contents

Advertisement

5.2.7

PC and printer

Interfaces
USB (Type A)
USB B
Ethernet
2. Select the required action.
3. Press Setup and complete the required settings.
4. Press OK to confirm the entry.
The DR 6000 features an Ethernet port and two USB ports on the
rear of the instrument, as well as one USB port on the front (refer
to
to send data and graphics to local and network printers, update
data and transfer data to a PC and within a network. These
interfaces can be used for the connection of a USB stick, an
external USB keyboard or a USB Barcode handset scanner.
Note: Various accessories can also be connected simultaneously using
a USB hub.
A USB memory stick is used to update data, refer to
section 6.7.2, page
Note: A shielded USB cable must not be longer than 3 m!
Table 3 USB connector
Description
This USB interface is for the connection of a printer, of a USB storage device or of a USB
keyboard.
This USB port is only intended for the connection of the DR 6000 to a PC (with installation of the
corresponding software).
The Ethernet port is intended for data transfer to a PC without installed software or in a local
network . Only use a shielded cable (e.g. STP, FTP, S/FTP) with a maximum length of 20 m for
the Ethernet port.
1. Press PC & Printer in "Instrument Setup".
The instrument then returns to "Instrument Setup".
Figure 1
and
Figure 2, Page
116.
A list with information about the connections opens.
Standard programs
19). These interfaces can be used