Single Wavelength (Absorbance, Concentration And Transmittance Readings); Set Up Single Wavelength Mode - Hach DR 6000 User Manual

Laboratory spectrophotometer for water testing
Hide thumbs Also See for DR 6000:
Table of Contents

Advertisement

Expanded programs
6.3
Single Wavelength (absorbance, concentration and transmittance
readings)
6.3.1

Set up Single Wavelength mode

Options
More
Recall measurement
data symbol
% Trans/Abs
Wavelength
Timer icon
100
4. Press OK to confirm.
The Single Wavelength mode can be used in three ways. For
sample readings at a single wavelength, the instrument can be
programmed to measure the absorbance, % transmittance or
concentration of the analyte.
1. Press Single Wavelength in the Main Menu.
2. Press Options for Parameter Setup.
Table 12 Options for Single Wavelength
Description
For further Options
Recall saved measurement data, wavelength scans or time courses, refer to
page
44.
Switches from % transmittance to concentration or absorbance in reading mode.
To input the wavelength setting. Use the alphanumeric keypad to enter the reading wavelengths.
A wavelength in the range of 190–1100 nm can be entered.
This functions as a stopwatch. It helps to make sure that the steps of an analysis are correctly
timed (e. g. reaction times, wait times, etc., can be exactly specified). When the specified time
has elapsed, a sound is emitted. The use of the timer has no influence on the reading program.
Note: If a Favorite Program is deleted, it will stay in the User
Programs or Stored Programs.
Note: If the stored program is deleted in User Programs, it will also
be deleted in Favorites Programs.
Absorbance readings: The light absorbed by the sample is
measured in absorbance units.
% Transmittance measures the percent of the original light
that passes through the sample and reaches the detector.
Turning on the concentration factor allows selection of a
specific multiplier for converting absorbance readings to
concentration. In a graph of concentration versus the
absorbance, the concentration factor is the slope of the line.
section 5.3,