6.7.4.4

Dilution

6.7.4.5

Performing AQA measurement

6. Enter the number of measurements for multiple definitions.
Press OK to confirm.
7. Enter the tolerance as a percentage. Press OK to confirm.
8. Press OK to confirm all the entries.
1. Press Dilution.
2. Press On.
3. Select the required interval. Press OK to confirm the entry.
4. Select a unit for the interval. Press OK to confirm the entry.
5. Enter the dilution factor. Press OK to confirm.
6. Enter the tolerance as a percentage. Press OK to confirm.
7. Press OK to confirm all the entries.
If a reminder function has been set, once the set interval has
elapsed the requirement to perform an AQA measurement is
issued.
1. Select the required AQA measure – for example a standard
measurement here.
2. Perform the standard measurement via stored program or
barcode program.
3. An AQA symbol is displayed in the measurement window,
the AQA key in the toolbar is displayed as yellow.
4. The AQA measure is shown in a log as Passed or Failed.
Expanded programs
125