while scanning to the right or to the left. The examples of the screen display for normal and
invert settings are shown in Figure 2-7. Set a desired setting to this parameter in accordance with
the procedure described in section 2.3.
Remarks
• The direction of the image data output to the printer (optional) follows the setting of
the display directcion . To output image data to the printer always in the same
direction, set the same setting to the display direction.
• While the screen display is inverted, the cursor keys
direction. (They work in the same direction as when
display direction.)
Screen display direction [Normal]

2.3.3 disp. mode

You can specify the display mode by this parameter. The available settings are B and BA .
This product supports two display modes of B-mode (vertical cross section) and BA-mode (vertical
cross section and reflected waveform display). When setting B to this parameter, the scan result
is displayed in the B-mode. When setting BA to this parameter, the scan result is displayed in the
BA-mode. Set a desired setting to this parameter in accordance with the procedure described in
section 2.3.
Remarks
• When the display mode is switched between
data scanned in the B-mode is displayed in the BA-mode, and the data scanned in
the BA-mode is displayed in the B-mode. See Section 3.3.1 Mode switching.
Figure 2-7 Screen display inversion
- 17 -
▲▼
work in the opposite
normal
is selected for the
Screen display direction [invert]
B
and
BA
after scanning, the